- Neeo S available in different models: with stationary tube and Image Intensifier 9” or 12”
- Horizontal run 210 mm
- Arc orbital Movement 130°
- WIG-WAG ±12°
- C-arm depth 659 mm
- S. I. D. 730 mm
- Arm rotation around horizontal axis ±270°
- Generator power 3,5kW
- Generator operating frequency 40 kHz
- KV range 40 -h 110 kV
- Max current in fluoroscopy 5 mA - 7 mA «pulsed» – 10 mA One Shot
- Max current in «Boosted» 12 mA
- Max current in radiography 65 mA
- Maximum Current Absorbed: In fluoroscopy: 6 A – 230 Vac Fluoro Pulsed - In Radiography: 22 A – 230 Vac Radio Pulsed
Operating Modes and Functionality
OPERATING MODALITIES OF MEMORY
CONTINUOS FLUOROSCOPY
PULSED FLUOROSCOPY
ONE SHOT FLUOROSCOPY
CONTINUOS , PULSED AND ONE SHOT LOW DOSE FLUOROSCOPY
RADIOGRAPHY WITH CASSETTE
Image Itensifer and CCD
- Useful Entrance Field Size: 215 mm / 160 mm / 120 mm
- Contrast Ratio on large area: 23:1, 25:1, 30:1
- Image resolution: 1000 x 1000 Pixels
- Digital Video Output: 12 bits
- Maximum frame speed: 30 fps
- MTF: Motorized Lens ≥ 50%
Software:
Functions:Fluoroscopy 25 images per second 1024x 1024 x 12 bit; Recursive filter and movement detection; Acquisition rate 1,3,6,12, 25 image/sec; Electronic rotation with 1° step; Horizontal and vertical inversion; Brightness and contrast; Maximum opacity fluoroscopy acquisition; Real time subtraction with auto/manual mask; Storage on disk up to 25 images per second; Virtual collimators; Grey scale inversion; Spatial filters (edge enhancement); Cineloop; Landmarking; Overview 4-9-16 images; Print on Windows USB2 printers; Storage examination on removable drive (USB2 Pen drive) in BMP
Measure tools (included in basic configuration): - Distance Calculator
- Angles
- Stenosis
- Overwrite text
- 12" and 9" image intensifier option available
- High Frequency monoblock X-ray generator 3.5 KW
- 1k CCD camera delivers sharp and detail-rich images
- Automatic” Technique (0 Points) with automatic control on kV and mA
- Electronic zoom selection according to the number of fields of the Image Intensifer
- Display touch screen (also working with work gloves) backlit and high contrast for the insertion of all parameters and for the display of any error messages or system issues
- Full Dicom(optional)
- foot-switch fluoroscopy control
- 210 mm horizontal C-arm run, with manual brake for locking
- Orbital movement 130°
- 270° on each side arm rotation, with manual brake for locking.
- 12° on each side C-arm swivelling with manual brake for locking.
- Monitor trolley with 2, 19 "high resolution monitors
- Safety:
- Filament current
- mA min and mA max
- Max Exposure Time
- Max Temperature of the monobloc
- Counting monoblock thermal units
- Timer with RX stop every 5 minutes (up to 99 minutes)
- max kV, min kV, max kV, max I
- Anode Rotation
- Microprocessor auto test
- DICOM package:
- DSA (Digital Subtraction Angiography)
- QA (Quantitative Analysis)
- HCF (High Contrast Fluoroscopy)
- Dicom BASE Verify and Storage
- Dicom PRINT Print on a DICOM Printer
- Dicom WORKLIST
- Dicom MPPS (Modality Performed Procedure Step)
- Dicom STCOMMIT (Storage Commitment)
- Dicom QUERY/RETRIEVE
- Dicom CD/DVD
- RTP Real Time Processing NW
- Dosimetric system
- Stainless steel cassette holder
- Cross laser locator
- Sony Thermal printer A4 format or A6 format
- Sterile drapes